Page 94 • MNKS025 ║ Issue/Rev. 0.1 (11/18)
7. Verify the connection parameters only allow the display to connect with devices that
are part of the Legally Relevant system.
• Select Reconnect
• Verify that the serial number written down earlier is shown in the upper left
hand corner of the meter display overview screen.
• Note that you can also select the Meter Information Screen and the serial
number is displayed in the General category of that page.
8. For a remote display, inspect the connections to the display; verify there are
NO connections to the display except for power (CN4) and Ethernet (ETH1).
Any other connections to the display are not allowed as they compromise
security. Bolt the enclosure closed and apply a seal wire such that it must be
destroyed to open the enclosure again.
Primary URL Connection Method
1. Ensure that only the UMCB and Display are connected together and that no
other Ethernet devices are active. If the UMCB and Display are in the same
enclosure, ensure the display is the only device connected to the Ethernet ports
on the UMCB. Note that this process assumes that the URL of the meter has been
set and is known.
2. Configure the display.
• Select Configuration
• Select Security Level 5
MPU Series C I/O/M Manual
Integrated Touch Screen Display
Содержание Smith Meter MPU 1600c
Страница 10: ...Page 10 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 34: ...Page 34 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 38: ...Page 38 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 98: ...Page 98 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 106: ...Page 106 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 112: ...Page 112 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 119: ...Issue Rev 0 1 11 18 MNKS025 Page 119 Page intentionally left blank ...