Issue/Rev. 0.1 (11/18) ║ MNKS025 • Page 27
MPU Series C I/O/M Manual
UMCB – Ultrasonic Meter Control Board
4 – UMCB (Ultrasonic Meter Control Board)
NOTE:
On the UMCB cover shield legends position 1 is indicated by a circle enclosed
in a square.
4.1.
Wiring Examples/Information
4.1.1.
Power Supply
The electronics are designed to be powered by 24 VDC. See specification sheets for
additional information.
The shield of the cable should be connected only at the electronic enclosure, preferably
via the cable gland.
Connect input power cable to connector CN1.
DC Input Wire
Terminal
24 VDC (+)
CN1-1 (or CN1-2)
Ground
CN1-3 (or CN1-4)
4.1.2.
Input / Output Wiring
•
Digital I/O
- (QTY 2) Digital Inputs (CN2)
- (QTY 2) Digital Pulse Outputs/Digital outputs (CN3)
- Alarm Output (CN3)
•
Analog I/O
- (QTY 2) 4 to 20 mA Input (CN6)
- 4 to 20 mA Output (CN5)
•
Communications
- (QTY 2) ANSI/IEEE 802.3 Ethernet Port (ETH1 and ETH3)
- Fiber Optic Port - 100Base-FX (ETH2)
- RS-485 Half-Duplex (2-wire) (CN4)
Содержание Smith Meter MPU 1600c
Страница 10: ...Page 10 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 34: ...Page 34 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 38: ...Page 38 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 98: ...Page 98 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 106: ...Page 106 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 112: ...Page 112 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 119: ...Issue Rev 0 1 11 18 MNKS025 Page 119 Page intentionally left blank ...