Page 114 • MNKS025 ║ Issue/Rev. 0.1 (11/18)
MPU Series C I/O/M Manual
Appendix A – Description of Legally Relevant Parameters
12 – Appendix – Description of Legally Relevant Parameters
MPU Series Meter Database Parameters
The following is a list of all parameters. All parameters at security levels 3, 4 and 5 (Commissioning, Owner and
Weights and Measures) are Legally Relevant; there are no Legally Relevant parameters at levels 1 and 2.
Parameter
Name
Menu
Range
Nominal
Description
Electronic
Sealing Mode
Modes
-no alarm
-OIML
Default is “no alarm”
Electronic seal mode; determines
meter behavior on sealed and
unsealed conditions.
Length
Meter Body
50 mm
1500 mm
Depends on meter size
Length measured between transducer
faces (one parameter for each path)
Angle
Meter Body
40 degrees
60 degrees
and
-40 degrees
-60 degrees
Depends on path configuration of
meter spool
Angle between transducer path and
flow direction (one parameter for each
path)
Lateral Position
Meter Body
-1.0
1.0
Depends on path configuration of
meter spool
Lateral level of each path:
-1=bottom, zero=center, 1=top (one
parameter for each path)
Upstream Port
Number
Transducer
0
15
Depends on path configuration of
meter spool
Upstream transducer electrical port
number. (one parameter for each path)
Downstream
Port Number
Transducer
0
15
Depends on path configuration of
meter spool
Downstream transducer electrical port
number. (one parameter for each path)
Initial Profile
Factor
Factory
Settings
0.8
1.0
Depends on path configuration of
meter spool
Profile (reference factor) to use before
the learner runs. (one parameter for
each path)
Disable
Modes
0 or 1
0 (enabled)
Disable this path (will be substituted).
(one parameter for each path)
Transducer
Delay
Transducer
-3.0 us
+15.0 us
Depends on transducer type
Transducer delay value (node 1 if
multiple nodes are programmed). (one
parameter for each path)
Transducer
Delay (node 2)
Transducer
-3.0 us
+15.0 us
Depends on transducer type
Transducer delay value (node 2). (one
parameter for each path)
Transducer
Delay (node 3)
Transducer
-3.0 us
+15.0 us
Depends on transducer type
Transducer delay value (node 3). (one
parameter for each path)
Internal
Diameter
Meter Body
70 mm
1500 mm
Depends on meter size
Average spool internal diameter at
reference temperature.
Wall Thickness
Meter Body
20 mm
100 mm
Depends on meter construction
Average thickness of the spool wall;
used for p/t correction of spool
dimensions.
Material
Meter Body
Carbon steel,
Stainless steel,
Duplex
Depends on meter construction
Material of spool construction; used for
p/t correction of spool dimensions.
Reference
Temperature
Meter Body
15 C
30 C
Depends on ambient conditions
Temperature of the spool when the
path lengths and inner diameter were
measured.
Содержание Smith Meter MPU 1600c
Страница 10: ...Page 10 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 34: ...Page 34 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 38: ...Page 38 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 98: ...Page 98 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 106: ...Page 106 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 112: ...Page 112 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Страница 119: ...Issue Rev 0 1 11 18 MNKS025 Page 119 Page intentionally left blank ...