Calibrating the Bar Code Reader
To calibrate the bar code reader, do the following:
1. Insert a cartridge in
Slot 2
of the SLR Autoloader. Make sure the cartridge has a
bar code label that is:
•
Tandberg approved for SLR Autoloader
•
Clean
•
Properly positioned on the cartridge.
2. Power up the Autoloader.
3. Press and hold down the
Menu
button while you press and release the
Load
button.
4. Continue holding down the
Menu
button until the READY light turns off. This
will take abt. 10 seconds.
5. From the DIAGNOSTICS menu, scroll down to and select Barcode Calibration.
The SLR Autoloader fetches the cartridge from Slot 2 and moves the cartridge
back and forth across the bar code reader sensor.
It displays the upper and lower values as it progresses through the calibration
and then displays one of the following:
•
If the calibration passes
PASSED
•
If the calibration fails or is aborted
ABORTED
•
If there is no cartridge in Slot 2
NO CARTRIDGE IN SLOT 2
6. If there is no cartridge in Slot 2, exit the DIAGNOSTICS MENU, open the door,
place a cartridge in Slot 2, reinsert the magazine and start the procedure again
from Step 3.
Tandberg SLR Autoloader Service and Repair Manual
11