6. Attach the thermocouple to the stage as
follows:
a. Position the tip of the thermocouple at
approximately one half the length of the
samples you will be measuring. The tip
should be angled toward the sample
location.
After the thermocouple is installed, its tip may
interfere with the zero length operation. Tempo-
rarily move the thermocouple when performing
this operation.
b. Hold the thermocouple against the stage
assembly, and put on the spring clip to
hold the thermocouple in place.
7. Replace the stage shield on the base of the
stage.
8. Move the furnace back into position and
lower it. See Figures 4.4 and 4.7 for the
correct probe setup with fiber and film
samples.
Calculating
the Offset
When you use the film/fiber accessory, analysis
must compensate for the thickness of the stage
platform and the upper portion of the probe.
Perform the Auto zero offset procedure to
compensate for the thickness. (See the instru-
ment control software online documentation for
further instructions on performing this operation.)
Film/Fiber Accessory
NOTE:
Содержание TMA 2940 CE
Страница 20: ...TA INSTRUMENTS TMA 2940 CE XX...
Страница 22: ...Introduction 1 2 TA INSTRUMENTS TMA 2940 CE...
Страница 36: ...Introduction 1 16 TA INSTRUMENTS TMA 2940 CE...
Страница 38: ...Installation TA INSTRUMENTS TMA 2940 CE 2 2...
Страница 136: ...Using Your Options TA INSTRUMENTS TMA 2940 CE 4 48...
Страница 138: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 2...
Страница 148: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 12...
Страница 150: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 CE 6 2...
Страница 178: ...Index TA INSTRUMENTS TMA 2940 CE I 8...