Using Your Options
TA I
NSTRUMENTS
TMA 2940 CE
4–10
The Auto Measure Offset does not need to be
changed unless you:
Reset the instrument parameters
Use a new film/fiber probe and stage
assembly.
Change the Auto Measure Offset as follows:
1. Record the value listed as Auto Measure
Offset from the TA controller.
2. Use calipers to measure the gap in the Auto
Measure Offset Gauge, and record this
value as the
Actual Size
.
3. Open the probe and insert the gauge as
shown in Figure 4.2.
Figure 4.2
Inserting the
Auto Measure Gauge