TMA 2940 CE
2–21
Figure 2.10
Address Selector Dial
(Showing an Address of 5)
If you change the address after the TMA is
powered on, you must press the TMA Reset
button to enter the new address. Wait 30 sec-
onds after releasing the Reset button, the green
Ready light should begin to glow steadily.
Reconfigure the instrument with the controller to
bring the instrument back online.
The instrument’s GPIB address is displayed
during startup and can also be viewed on the
instrument’s status display.
NOTE:
Installing the Instrument