Introduction
1–10
TA I
NSTRUMENTS
TMA 2940 CE
Table 1.1
TMA 2940 CE Keypad
Functions
(continued)
Key/Function
Explanation
STOP
If an experiment is
running, this key ends the
method normally, as
though it had run to
completion;
i.e.
, the
method-end conditionsgo
into effect, and the data
that has been generated is
saved. This is the same
function as
Stop
on the
controller.
If an experiment is not
running (the instrument is
in a stand-by or method-
end state), the
STOP
key
will halt any activity (air
cool, all mechanical
motion, etc.).
REJECT
If an experiment is
running,
SCROLL
-
STOP
ends the method normal-
(Hold down
ly, as though it had run to
SCROLL
and
completion;
i.e.
, the
press
STOP
)
method-end conditions go
into effect, and the data
that has been generated is
discarded
. This is the
same function as
Reject
on the controller.
(table continued)
Содержание TMA 2940 CE