Maintenance and Diagnostics
TA I
NSTRUMENTS
TMA 2940 CE
6–12
Fuse F1 is in the circuit between the POWER
switch and the instrument control circuits. All
power for internal operations and instrument
functions, except heater power, passes through
this fuse. If this fuse blows, you will get no
response from the system.
Fuse F2 is in the circuit between the main
electrical input and the
POWER
switch. This
fuse protects all internal components, including
the furnace. If this fuse blows, you will get no
response from the instrument when you attempt
to turn it on.
Furnace Power Check
Furnace power is always checked at the begin-
ning of a method. Power supplied to the furnace
is switched by a computer-controlled relay as
well as the
HEATER
switch located on the
instrument's front panel. The
HEATER
switch
must be on ( ) to start a method.
The light in the HEATER switch will glow only
after an experiment is initiated. The HEATER
switch will continue to glow, even if it is
switched to the OFF (0) position, until the "Stand
By" status code is displayed.
NOTE:
Содержание TMA 2940 CE
Страница 20: ...TA INSTRUMENTS TMA 2940 CE XX...
Страница 22: ...Introduction 1 2 TA INSTRUMENTS TMA 2940 CE...
Страница 36: ...Introduction 1 16 TA INSTRUMENTS TMA 2940 CE...
Страница 38: ...Installation TA INSTRUMENTS TMA 2940 CE 2 2...
Страница 136: ...Using Your Options TA INSTRUMENTS TMA 2940 CE 4 48...
Страница 138: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 2...
Страница 148: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 12...
Страница 150: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 CE 6 2...
Страница 178: ...Index TA INSTRUMENTS TMA 2940 CE I 8...