TA I
NSTRUMENTS
TMA 2940 CE
4–5
Film/Fiber Accessory
Operating
Procedures
Film or fiber samples, held in tension during
operation, are generally longer than expansion or
penetration samples; therefore, the small temp-
erature gradient normally present in the furnace
is more significant. A faster heating rate serves
to accent this gradient. For the best reproducibil-
ity, when creating a method, use a Ramp seg-
ment of 5
o
C/min or less for your heating rate.
(See the instrument control software online
documentation for details.)
When comparing run data, make sure that all
comparison scans are made with the thermo-
couple in the same position and at the same
heating rate.
Accessory
Installation
Because the configuration of the film/fiber
accessory is different from that of the other
types of probes, you need to perform the fol-
lowing general steps to install the special film/
fiber probe and stage:
Remove the existing probe
Remove the stage
Change the instrument mode
Install the film/fiber probe
Install the film/fiber stage
Calibrate the probe
Calibrate temperature and cell constant.
NOTE:
Содержание TMA 2940 CE
Страница 20: ...TA INSTRUMENTS TMA 2940 CE XX...
Страница 22: ...Introduction 1 2 TA INSTRUMENTS TMA 2940 CE...
Страница 36: ...Introduction 1 16 TA INSTRUMENTS TMA 2940 CE...
Страница 38: ...Installation TA INSTRUMENTS TMA 2940 CE 2 2...
Страница 136: ...Using Your Options TA INSTRUMENTS TMA 2940 CE 4 48...
Страница 138: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 2...
Страница 148: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 12...
Страница 150: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 CE 6 2...
Страница 178: ...Index TA INSTRUMENTS TMA 2940 CE I 8...