Maintenance and Diagnostics
TA I
NSTRUMENTS
TMA 2940 CE
6–8
Replacements
Occasionally, you may need to replace a broken
or worn-out part of the TMA. Any replacements
needed, other than those discussed in this
manual, must be supplied and installed by
qualified TA Instruments service personnel. Call
(302) 427-4050 for service.
Replacing the
Thermocouple
You may find that the need arises for a new
thermocouple (PN 944344.902) as a result of
normal wear and tear, accidental breakage,
contamination of the thermocouple, etc.
Before removing the thermocouple, you will need
to perform the following steps:
1. Take off the TMA balance enclosure, the
stage shield, and the spring clip holding the
thermocouple in place.
2. Thread the thermocouple carefully down
through the opening in the platform, and
unplug it.
Replace the sample thermocouple, following
the remainder of the instructions.
3. Insert the two-pin, color-coded thermocouple
connector into the fitting as shown in the
figure on the next page.