Installation
TA I
NSTRUMENTS
TMA 2940 CE
2–10
Inspecting
the System
When you receive your TMA, look over the
instrument and shipping container carefully for
signs of shipping damage, and check the parts
received against the enclosed shipping list.
If the instrument is damaged, notify the carrier
and TA Instruments immediately.
If the instrument is intact but parts are missing,
contact TA Instruments.
A list of TA Instruments offices can be found in
Appendix A of this manual.
Drying out the instrument may be needed, if it
has been exposed to humid conditions. Certain
ceramic materials used in this equipment may
absorb moisure, causing leakage currents to
exceed those specified in the applicable stan-
dards (see page xi) until the moisture is elimi-
nated. It is important to be certain that the
instrument ground is adequately connected to
the facilities ground for safe operation.
Run the following method to dry out the instru-
ment (refer to “Running Experiments” for
further information).
1
Ramp at 10°C/min to 400°C
2
Isothermal for 30 min.
NOTE:
Содержание TMA 2940 CE
Страница 20: ...TA INSTRUMENTS TMA 2940 CE XX...
Страница 22: ...Introduction 1 2 TA INSTRUMENTS TMA 2940 CE...
Страница 36: ...Introduction 1 16 TA INSTRUMENTS TMA 2940 CE...
Страница 38: ...Installation TA INSTRUMENTS TMA 2940 CE 2 2...
Страница 136: ...Using Your Options TA INSTRUMENTS TMA 2940 CE 4 48...
Страница 138: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 2...
Страница 148: ...Technical Reference TA INSTRUMENTS TMA 2940 CE 5 12...
Страница 150: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 CE 6 2...
Страница 178: ...Index TA INSTRUMENTS TMA 2940 CE I 8...