Operating Manual
5
1300-87076-1A Rev. 1
1.2.3 Electrical Installation
THIS INSTRUMENT MUST BE GROUNDED
The appropriate power supply and cord combination should be connected to the equipment BEFORE
connection to the mains supply.
Before connection, please ensure that the electrical supply is suitable for the power requirements below:
Current
Current
Voltage
Voltage
1.0 A
100 V to 230 V ~ 50/60 Hz
The equipment is provided with a power supply and a selection of four mains cords, consisting of:
- Type B (US) for 120 V a.c. installations
- BS1363 style, 3-pin (UK), Type F (EU) and Type I (Aus / China) for 230 V a.c. installations
Select the power cord appropriate for your electrical installation and discard the others.
Mains Cord Replacement
Should one of the power cords not be suitable for connecting to the power supply, replace the plug with a
suitable alternative.
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN
NOTE
: Refer to the equipment specification to ensure that the plug and fusing are suitable for the voltage and
current stated. The cores in the mains cord are as follows:
230 V a.c.
230 V a.c.
120 V a.c.
120 V a.c.
BROWN
LIVE/HOT
BLACK
BLUE
NEUTRAL
WHITE
GREEN/YELLOW
EARTH
GREEN
Should the mains cord require replacement, a cable of of harmonized code H05VV-F with a cross section of
0.75 mm² (18 AWG) must be used. The IEC C13 connector (for connecting to the equipment power supply) must
be suitable for the supply voltage and current of the equipment.