Operating Manual
18
1300-87076-1A Rev. 1
5.1.2 Table Tab Page
The table tab page will show all the data results. If the user needs to know more details of one data item, tap the
check box column to select the data. The detail will be shown on the
Data
page and the
Graph
page.
Figure 13. Table Tab Page
Tap the
Sample
column to change the sample name. The
Blank
can also be renamed, but this is NOT
suggested. You need to measure the
Blank
again if the original blank data has been renamed.
Note
: Only one data item can be checked at one time.
5.1.3 Graph Tab Page
The graph of measuring samples can be enlarged or minimized. The axis can be moved by dragging the graph.
Long touch the x- or y-axis on the graph to turn the modified graph back to default.
Figure 14. Graph Tab Page
5.2 Protocol Operation
1. On the main menu, tap on the
Protein A280
icon to enter the protocol section.
2. Ensure the sample window and the cover window on the detection arm are clean.
3. Select the correct
Method
type according to the experiment. The default selection is BSA.