Operating Manual
32
1300-87076-1A Rev. 1
On the Sample Data tab page of Std. Curve method protocol (Figure 27), the data information parts have the
features below (Table 18).
The function icons will be activated after finishing the protocol setting on the Custom Setting tab page.
Figure 27. Sample Data Tab Page
Table 18. Sample data tab page information.
Features
Description
[conc.]
The concentration is calculated from absorbance at user-defined wavelength. The unit
is also user-defined but only shown on the custom setting page. This is an optional
function.
Abs
Displays the absorbance at user-defined wavelength which is normalized to a 10 mm
pathlength equivalent.
λ (Anlaysis)
User-defined wavelength decided on the custom setting page.
λ (Corr.)
Self-defined wavelength for bichromatic normalization. This is an optional function and
the default is Off.
Name
The sample name can be inserted here. The default is Sample.
Pathlength
The light path chosen by the pathlength selector will be detected automatically and
length will be shown here.
The Standard tab page (Figure 28) is the page to measure standard samples absorbance at specific wavelength
and establish the standard curve. It only shows the standard data result. The features are shown below (Table 19).
Figure 28. Standard Tab Page
Function icons
Sample data information
Information tabs
Insert sample name
Selected pathlength
Curve type selection
Repetition frequency
Standard curve check box
Standard curve information
Function icons
Sample data information
Information tabs