Operating Manual
37
1300-87076-1A Rev. 1
2. Ensure the sample window and cover window on the detection arm are clean.
3. Insert required
Analysis Wavelength
. Spex NanoSNAP affords a maximum of 8 wavelengths.
4. Add appropriate solution of at least 1 µL and select the
Blank
icon to establish blank data.
5. Wipe away the blank solution off the sample window and the cover window.
6. Add your sample of at least 1 µL and select the
Measure
icon to enter sample measurement.
7. After the experiment, clean the sample window and the cover window on the detection arm with a lint-free
wipe. Use deionized water, ethanol or isopropanol if needed.
8. The correction function is optional, and can be turned on/off any time.
9. The default of the
Auto Run
function is off. If
Auto Run
is on, sample measurement will be performed
automatically after closing the detection arm