Operating Manual
43
1300-87076-1A Rev. 1
13.1 Error Messages
Message
Cause
Action
Er01
ERR_NO_SDCARD
Did not receive SD card signal in
1 second continuously
Return the unit for service
Er02
ERR_SELFTEST_NG
Automatically detected numerical
anomalies. Insufficient light source
intensity or excessive noise.
Return the unit for service
Er03
ERR_METER_NO_ANSWER
Optics module problem
Reboot the unit
Er04
ERR_METER_CALIBRATE
Optics module problem
Reboot the unit
Er05
ERR_UART_NO_ANSWER
Electronic module board problem
Reboot the unit
Er06
ERR_UART_WRONG_ANSWER
Electronic module board problem
Reboot the unit
Er07
ERR_UART_WRONG_COMMAND
Electronic module board problem
Reboot the unit
Er08
ERR_UART_TRANSMIT_OVERFLOW
Electronic module board problem
Reboot the unit
If the same error message appears after rebooting the unit, please return the unit for service.