Operating Manual
22
1300-87076-1A Rev. 1
Tap the
Sample
column to change the sample name. The
Blank
can also be renamed, but this is NOT
suggested. You need to measure the
Blank
again if the original blank has been renamed.
Note
: Only one data item can be checked at one time.
6.1.4 Graph Tab Page
In the protein assay protocol, there are 2 kinds of graphs: the standard curve graph and sample absorbance
graph.
They can be enlarged or minimized. The axis can be moved by dragging the graph. Long touch the x- or y-axis
on the graph to turn the modified graph back to default.
Figure 18. Graph Tab Page, Standard Curve View
6.2 Protocol Operation
1. On the main menu, tap on the
Protein Assay
icon to enter the protocol section.
2. Ensure the sample window and the cover window on the detection arm are clean.
3. Select the correct
Method
according to the protocol.
4. Add the appropriate solution of at least 1 µL and tap on the
Blank
icon to establish blank data. The screen
will automatically jump to the
Standard Curve
tab.
5. Wipe away the blank solution off the sample window and the cover window with a lint-free wipe.
6. Select the correct
Curve Type
according to your reagent.
Note:
1. Linear curve type requires minimum 2 DIFFERENT concentrations to establish standard curve.
2. Interpolation curve type requires minimum 2 DIFFERENT concentrations to establish standard curve.
3.
2
nd
order polynomial curve type requires minimum 3 DIFFERENT concentrations to establish standard
curve.