SPELLMAN HIGH VOLTAGE
Introduction 1
Installation and Operating Manual
Page1-4
118149-001 REV F
Line Voltage
24V
DC
±1V
DC,
2.5A
(20W option). 4A (50W option)
The following table defines the power line requirements for the generators.
NOTE: THE FOLLOWING TABLE CONTAINS RECOMMENDED VALUES FOR THE 24V INPUT POWER.
A POOR QUALITY INPUT LINE MAY RESULT IN THE INSTALLER HAVING TO
DERATE
THE
GENERATOR'S
MAXIMUM
POWER
Mains
Voltage
Minimum
Recommended
Minimum
Recommended
Ground Wire Size
Apparent
Mains
Resistance
24VDC
J1-1
J1-2
J1-3
#20 AWG (0.52 mm
2
)
#20 AWG (0.52 mm
2
)
#20 AWG (0.52 mm
2
)
0.033
0.033
0.033
24V RETURN
J1-5
J1-6
J1-7
#20 AWG (0.52 mm2)
#20 AWG (0.52 mm2)
#20 AWG (0.52 mm2)
0.033
0.033
0.033
1.2.4
Environment Requirements
Operating Environment
Operating Temperature
0 to 40
C (32 to 104
F).
Relative Humidity
10 to 95%, non-condensing.
Atmospheric pressure range
500 to 1060 hPa (375 to 795 mm Hg).
TRANSPORT AND STORAGE
Ambient temperature range
-20 to 70
C (-4 to 158
F).
Relative humidity
5 to 95%, non-condensing.
Atmospheric pressure range
500 to 1060 hPa (375 to 795 mm Hg).