SPELLMAN HIGH VOLTAGE
Introduction 1
Installation and Operating Manual
Page1-3
118149-001 REV F
1.1
XRB011 DESCRIPTION
The XRB011 MONOBLOCK
is a complete integrated system consisting of a high voltage power supply
(HVPS), filament supply, X-ray tube and oil encapsulant which provide the required high voltage insulation in
one compact enclosure. The combination of proprietary control system and protection circuitry enables the
supplies to operate under arcing and extreme transient conditions without damage or interruptions. Additional
advantages are the elimination of high voltage cables and extremely low leakage X-ray radiation.
The XRB011 incorporates local and remote programming, monitoring, and fault indicators including safety
interlock. The X-ray source is a sealed unit containing a HVPS and an X-ray tube. The insulating oil provides
electrical insulation for the high voltage sections of the power supply and the X-ray tube in a sealed tank. The
oil also functions as a coolant to carry heat away from the tube. Convection cooling augmented by customer
provided minimum 50cfm external fan is required for the 50W option. A rubber bellows in the tank of the X-ray
source compensates for the expansion of the oil as the oil temperature varies with operating conditions thereby
eliminating the need for bulky overflow tank.
.
1.2
TECHNICAL SPECIFICATIONS
1.2.1
Generator Hardware Specifications
kVp range:
35 to 80 kVp output capability
kVp steps:
Continuous with 12 bits resolution
kVp accuracy:
<=1% (measured after kVp rises to the
peak level)
Ripple (kV):
<= 1% Peak to Peak
Settling time:
< 10ms to within 95% of the
programmed voltage
Reproducibility:
<0.5%
Stability:
<=0.01% per 8 hours after a ½ hour
warm up
Temperature Coefficient:
<=100ppm/
℃
Time range:
XRB011 is specified as a Continuous
operation.
mA :
250µA maximum for 20W option
700µA maximum for 50W option
mA accuracy:
<2.5% (measured after mA rises to
stable DC level)
mA range
0 to 250 µA for 20W option
0 to 700 µA for 50W option
Reproducibility:
<0.5%
1.2.2
Generator Control Modes / Application Features
Manual Operating Mode
– 2 Parameters Mode (kV, mA)
System Communication Protocol / Fault & Error Management
See details in the Serial Communication Protocol Specifications
1.2.3
Power Supply Requirements
Single Phase