SPELLMAN HIGH VOLTAGE
Installation 2
Installation and Operating Manual
– XRB011
Page 2-14
118149-001 REV F
The interlock, serial communication and exposure buttons interface shall be made available via a multi
signal cable. Refer to *-* for a schematic of isolation and signal direction.
2.5.1
Pin Layout of the Multi interface
The multi signal interface shall have a female 25 pin D-Sub (J1). Twisted pairs shall be used where
applicable. The pin layout is given in Table 2-3.
PIN
SIGNAL
PARAMETERS
1
+24V
+24Vdc±1Vdc @ 4A
2
+24V
+24Vdc±1Vdc @ 4A
3
+24V
+24Vdc±1Vdc @ 4A
4
NC
No connection
5
+24V RETURN
+24V RETURN
6
+24V RETURN
+24V RETURN
7
+24V RETURN
+24V RETURN
8
Signal Ground
Signal Ground
9
Interlock Input
Input, Active low, Interlock is low safe to enable high
voltage. Connect to +24V Return
10
kV Monitor
Output, 0 to 8V = 0 to rate
d output voltage. Zout=100Ω
11
µA Monitor
Output, 0 to 10V = 0 to rated output current. Zout=100Ω
12
X-Ray Ready status
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
13
X-Ray ON status
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
14
Filament Standby status
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
15
Over Voltage Fault
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
16
Over Current Fault
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
17
ARC Fault
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
18
Filament Current Limit
Fault
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
19
Signal Ground
Signal Ground
20
Interlock Output
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
21
µA Program
Input,
0 to 10V = 0 to rated output current. Zin=10kΩ
22
kV Program
Input, 0 to 8V = 0 to rated output voltage. Zin=10kΩ
23
X-Ray ON Command
Input, Active low,
Low (short)=X-Ray ON
High (open)=X-Ray OFF
Internal pull up resistor to +15V
24
Signal Ground
Signal Ground
25
Over Temp.
Output, Active Low,
Open Collector, 24Vdc @ 10mA max
Table 2-3 Pin layout of multi signal connector
Refer to chapter 3 for detailed requirements and function descriptions.
2.5.2
GROUND CONNECTION
NOTE: THE INSTALLER SHOULD ENSURE THAT ALL CABLE CONNECTIONS TO THE
GENERATOR ARE SECURE, AND ALL CABLES EXTERNAL TO THE GENERATOR
ARE ADEQUATELY PROTECTED AGAINST ACCIDENTAL DISCONNECTION.