SPELLMAN HIGH VOLTAGE
Interfacing
3
Installation and Operating Manual
– XRB011
Page 3-19
118149-001 REV F
X-RAY READY STATUS:
Indicates that there are no faults and the interlock is closed. The status is
indicated by NO FAULT (X-RAY READY) via
RS-232 or Ethernet as (ARG 0). Analog output signal (J1-12) is
active low.
X-RAY ON STATUS:
Indicates that the X-RAY is ON or OFF. This status is indicated by via RS-232or
Ethernet as X-RAY ON (ARG 1), X-RAY OFF (ARG 0).
Analog output signal (J1-13) is active low (Low X-
RAY is on, High=X-RAY is off).
OVERVOLTAGE FAULT:
Indicates the over voltage protection circuitry has caused the high voltage to turn off.
Over voltage protection is internally set to 82kV. This fault is indicated by High kV via RS-232 or Ethernet as (ARG
6), Over Voltage. Analog output signal (J1-15) is active low.
OVER CURRENT FAULT:
Indicates the output current has exceeded 275µA (20W option), 710µA (50W option)
or if the allowable percentage of error between actual and programmed emission currents is exceeded resulting in the
HV to be turned off. This fault is indicated by High mA via RS-232or Ethernet as (ARG 3). Analog output signal
(J1-16) is active low.
ARC FAULT
: Indicates that an arc has occurred. Occurrences of one arc will shutdown the high voltage and
latched. This fault is indicated by ARC FAULT via RS-232 or Ethernet as (ARG 2). Analog output signal (J1-17)
active low.
UNDER VOLTAGE FAULT:
Indicates a failure in the voltage regulation circuitry less than <35kV. This fault
occurs when there is a lack of output power to maintain regulation and will result in shutdown of the HV. This fault is
indicated by via RS-232 or Ethernet as (ARG 4). There is no analog output signal.
FILAMENT CURRENT LIMIT FAULT:
Indicates the filament current exceeded the safe operating current of the
X-Ray tube. This fault is indicated by FILAMENT LIMIT via RS-232 or Ethernet as (ARG 10). Analog output
signal (J1- 18) is active low.
FILAMENT STANDBY STATUS:
Indicates the X-RAY is off and the filament current is in standby mode. This
status is indicated by FILAMENT STANDBY via RS-232 or Ethernet as (ARG 11). Analog output signal (J1-14) is
active low.
WATCHDOG TIMER:
Indicates the host computer has lost communication and with the HVPS system for a period
greater than ten second. This feature is enabled via RS-232 host command. This fault is indicated via RS-232 as
(ARG 7) Watchdog Time- out. See digital manual for details.
OVER TEMPERATURE:
Indicates that the internal oil temperature has exceeded 65 degree C.
3.10
REMOTE MODE (DIGITAL CONTROL)
G.U.I Installation software will be provided up on request.
Start the G.U.I.
Read agreement then click AGREES...
Содержание XRB011
Страница 11: ...SPELLMAN HIGH VOLTAGE Introduction 1 Installation and Operating Manual Page1 6 118149 001 REV F 50W OPTION ...
Страница 69: ...118150 001 REV B Page 33 of 66 Figure 9 Web Page 1 Contact Information ...
Страница 74: ...118150 001 REV B Page 38 of 66 4 3 2 7 Version Information Figure 15 Version Information ...