Bellows Vent holes:
Do not block or insert anything into the vent holes located on top cover as shown. These two holes provide venting
for the internal bellows that is used to compensate for the expansion of the oil as the oil temperature varies.
Do not block any vents holes on the cover
Cooling:
A customer supplied minimum 50 CFM fan should be used for the 50W option to maintain safe operating
temperature for MONOBLOCK
X-ray generator. The air flow should be direct at the heat sink on the side of the
unit and at the side of the control board compartment. During operation the internal oil temperature should be below
60C and should not exceed 65C. See below Figure for fan location.
FAN