118150-001 REV. B
Page 22 of 66
3.4.5.10 Set Ramp Time <29>
Description:
It sets the KV and uA ramp time to full scale value. Units are milli-
seconds. Recommended range is 1 to 1000 milliseconds. Default value is
250 milliseconds. This command is password protected and the
userconfig command must be sent first.
Direction:
Host to supply
Syntax:
<STX><29><,><ARG><,><CS><ETX>
Where <ARG> = 1-1000 milliseconds
Response:
<STX><29><,><ARG><,><CS><ETX>
Where:
<ARG>= $ or error code.
Error Codes: 1 = receive error, 2 = unrecognized command