118150-001 REV. B
Page 12 of 66
3.4.4 RESPONSE OVERVIEW
The command responses will follow the same format as outlined above in
section 2.1. This list is comprised of Commands with complex responses
only. Commands using a simple response will use the <$> character
(ASCII 0x24) as a “Success” response or a single character error code.
These responses will be eight ASCII characters in length.
Response Name
<CMD>
<ARG>
Request KV
Setpoint
14
1-3 ASCII
Request MA
Setpoint
15
1-3 ASCII
Request Status
22
3 ASCII
Request DSP
Software Version
23
11 ASCII
Request Model
number
26
5 ASCII
Get KV monitor
60
1-3 ASCII
Get MA monitor
61
1-3 ASCII
Get X-RAY Status
98
1 ASCII