SPELLMAN HIGH VOLTAGE
Trouble Shooting
4
Installation and Operating Manual
– XRB011
Page 4-26
118149-001 REV F
TABLE 2
– GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS – FOR
ALL ME EQUIPMENT AND ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The XRB011 is intended for use in the electromagnetic environment specified below. The customer or the
user of the XRB011 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for input/output
lines
±
2 kV for power
supply lines
±
1kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage
dips,
short
interruptions
and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the XRB011 requires
continued operation during power mains
interruptions, it is recommended that the
XRB011 be powered from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical
commercial
or
hospital
environment.
NOTE
U
T is the A.C. mains voltage prior to application of the test level.
TABLE 3
– GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY – FOR ME
EQUIPMENT AND ME SYSTEMS THAT ARE NOT LIFE-SUPPORING
Guidance and manufacturer’s declaration – electromagnetic immunity
The XRB011 is intended for use in the electromagnetic environment specified below. The customer or the user of the
XRB011 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the XRB011,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Содержание XRB011
Страница 11: ...SPELLMAN HIGH VOLTAGE Introduction 1 Installation and Operating Manual Page1 6 118149 001 REV F 50W OPTION ...
Страница 69: ...118150 001 REV B Page 33 of 66 Figure 9 Web Page 1 Contact Information ...
Страница 74: ...118150 001 REV B Page 38 of 66 4 3 2 7 Version Information Figure 15 Version Information ...