118150-001 REV. B
Page 51 of 66
4.15 COMMANDS DESCRIPTION
4.15.1 Set KV <10>
Description:
The host requests that the firmware change the KV set point.
Direction:
Host to supply
Syntax:
<STX><10><,><ARG><,><ETX>
Where:
<ARG> = Value of KV in ASCII format. Units are tenths of KV. Example,
For 80KV, enter 800.
Example:
<STX>10,800,<ETX>
Response:
<STX><10><,><$><,><ETX>
or
<STX><10><,><ARG><,><ETX>
where <ARG> = error code
Error Codes: 1 = receive error, 2 = unrecognized command