SPELLMAN HIGH VOLTAGE
Installation 2
Installation and Operating Manual
– XRB011
Page 2-15
118149-001 REV F
2.5.3
CHASSIS GROUND
The chassis of the XRB011 MONOBLOCK must be grounded to the local earth ground and also to the
tube housing ground.
2.5.4
X-RAY PORT
ENSURE THAT THE EXIT PORT IS PROPERLY MATED TO COLLIMATOR OR SATISFACTORILY
SHIELDED WITH LEAD PLUG TO LIMIT EXPOSURE TO LEAKAGE RADIATION.
X-ray Safety Procedures must be followed when testing this unit. The XRB011 is capable of producing
Lethal Voltages and X-ray Radiation. Only proceed with operation of the HVPS after
Consulting with the Manufacturer and verification of X-Ray setup for the proper precautions.
Reading this entire document.
NEVER OPERATE THIS UNIT WITH AN OPEN X-RAY EXIT PORT.
It is recommended not to allow leakage radiation exceeding 0.5mR/hr at 5cm from any surface of the
MONOBLOCK.
2.6
FINAL CHECKS
The room interface connections may now be completed. Before power on, user needs to check the items
as below finally.
When finished all wiring, check that all connections are tight and secure.
Check that all cables are dressed neatly outside the cabinet, and secured as necessary.
Check the ground connection again.
THIS EQUIPMENT GENERATES DANGEROUS VOLTAGES THAT MAY BE FATAL.
PROPER GROUNDING OF ALL HIGH VOLTAGE EQUIPMENT IS ESSENTIAL.
WARNING X-RAY RADIATION EXPOSURE IS HAZARDOUS
Failure to follow these procedures may void the warranty.
Check the input voltage rating on the nameplate of the supply and make certain that this is the rating of the available
power source. Spellman MONOBLOCK
XRB011 operates on 24VDC±1VDC
WARNING
WARNING