6-21
FVS-1000 MMP1
6-6-2. 35 mm Integrating Sphere Offset/Gain
Adjustment
n
This adjustment procedure applicable to the LC-35 board
No. suffix -13 and higher.
Measuring instruments: Oscilloscope
PC with Hyper Terminal
Tool:
Connection cable for PC
Preparations
1.
Pull out the lamp house.
2.
Connect connection cable for PC from PC com1 port
to LC-35 board CN15 connector. (Refer to picture for
board lay out.)
Set the PC to Hyper Terminal and set the properties as
follows.
Baud Rate:
38400
Data bits:
8
Parity:
Even
Stop bits:
1
Flow Control: None
5
4
3
2
1
B
A
C
D
E
F
IC463
IC433
IC403
RV401
CN15
RV431
RV461
TP305
TP303
TP301
LC-35 board (Suffix -13 and higher, Side A)
6-6. Lamp House Adjustment