23. CHECKS AND ADJUSTMENTS
140
If the tilt angle shown on the display shifts from tilt angle 0° (zero point), the instrument is not correctly
levelled. This will adversely affect angle measurement.
Perform the following procedure to cancel the tilt zero point error.
PROCEDURE Checking and adjusting
1. Carefully level the SRX. If necessary, repeat the
procedures to check and adjust the bubble levels.
2. Select "Inst. cons." in <Configuration>
3. Select "Tilt offset".
23.3
Tilt Sensor
correction constant
in the X (horizontal)
direction
correction constant
in the Y (vertical)
direction