39
Checklist
Pump Cleaning and Functional Testing Checklist
The following checklist is only a guide to assist in establishing documentation of cleaning and functional
testing for the CADD
®
-Solis pump. The procedures are described in this Technical Manual. If service is
required, fill out a copy of this page and return it with the device.
Serial # _
Reference #
Date
I. Visual Inspection
☐ LCD
☐ Lock
☐ Remote dose jack
☐ Occlusion sensor seals
☐ Cassette sensors (3)
☐ Air detector
☐ Valves and expulsor
☐ Keypad
☐ Pump housing
☐ Cassette hinge area
☐ Indicator lights
☐ Battery door
☐ Latch handle
☐ AC power jack
☐ Battery compartment
☐ USB connector
II. Mechanical Inspection
☐ Keypad
☐ Cassette latch
☐ Battery Door
☐ Cassette lock
III. Functional Testing
☐ Power up, LCD
☐ Stop/Start pump, LED
☐ Air detector
☐ Latch/Lock
☐ Activation timing
☐ Battery fallout alarm
☐ Cassette sensor
☐ Remote dose cord
☐ Audible alarm
☐ Motor, reservoir is zero
☐ Doses given/attempted
IV. Occlusion Pressure Range Tests
Note:
Perform either downstream test 1 or test 2, but not both.
Downstream Test 1: Activations before alarm ________________
Downstream Test 2: High pressure alarm at _______________ psi
Upstream Occlusion Sensor Test: Pass __________ Fail ______
V. Accuracy Tests
Note:
Perform either the gravimetric or the volumetric test, but not both.
Gravimetric Accuracy Test
Pre
Delivery
Weight
(g)
Post
Delivery
Weight
(g)
Amount
Delivered
(mL)
Intended
Delivery
Volume
(mL)
Inaccuracy
Volume
(mL)
Accuracy
Error
Accuracy
Error
(%)
Volumetric Accuracy Test
Intended Delivery
Volume
(mL)
Actual Delivery
Volume
(mL)
Inaccuracy
Volume
(mL)
Accuracy
Error
Accuracy
Error
(%)