16
Pump Overview
Administrator Settings Specifications
KVO rate
•
0 mL/hr if continuous rate is 0 mL/hr
•
0.1 mL/hr if continuous rate > 0 mL/hr, with standard administration set
•
0.2 mL/hr if continuous rate > 0 mL/hr, with high volume administration set
Bolus interval type
•
Bolus interval
•
PCA lockout
Maximum delivery
rate (PCA dose,
clinician bolus,
intermittent bolus)
•
Model 2100: 250 mL/hr
•
Model 2110 with standard administration set: 250 mL/hr
•
Model 2110 with high volume administration set: 500 mL/hr
•
Max. delivery rate = continuous rate + bolus rate (PCA dose
or
clinician
bolus
or
intermittent bolus)
•
Boluses may not be delivered simultaneously
Delivery limit method
•
Delivery limit
•
Max doses per hour
•
Not in use
Delivery limit period
1 to 12 hours in increments of 1 hour
Pump stopped alarm
•
Informational
•
High priority
Res vol low trip point
1 to 999 mL in increments of 1 mL
Res vol empty alarm
•
Insistent and one time only
•
Non-insistent and repeating
Air detector
•
On
•
Off
Air detector sensitivity
Low Sensitivity:
Single bubble > 400 μL
High Sensitivity:
Single bubble > 150 μL
Upstream occlusion
sensor
•
On
•
Off
Note:
The upstream occlusion sensor is automatically disabled during use
with medication cassette reservoirs.
Downstream
occlusion sensitivity
High Sensitivity:
When the high pressure alarm threshold is reached, the
downstream occlusion alarm is triggered immediately.
Low Sensitivity:
When the high pressure alarm threshold is reached, the
downstream occlusion alarm is delayed for 2 seconds. This allows for the
pressure to stabilize before a possible alarm. If the pressure stabilizes below
the high pressure alarm threshold before the 2 second delay is complete,
the alarm will not occur.
PM reminder
Interval: 1 to 24 months in 1 month increments
Enable: On or Off
Custom keypad code
001 to 899 in increments of 1
Custom clinician code 001 to 899 in increments of 1
Custom admin. code
001 to 899 in increments of 1
Time format
•
00:00 to 23:59 military
•
12-hour
am
/
pm
Date format
•
US standard (month/day/year)
•
European standard (day/month/year)
•
International standard ISO 8601:2004 (year/month/day)
Alarm volume
•
High
•
Medium
•
Low
Key beep
•
On
•
Off