31
Annual Inspection and Testing
☐
Doses Given and Doses Attempted Check
36. Stop the pump by pressing stop/start
9
, then select
Yes.
37. From the View Reports menu, select
Given and PCA Dose Counters.
38. The display should show “Total Given 1 mL” and “PCA Doses Given 1, PCA Doses
Attempted 3.” If the steps in the Remote Dose Cord Check have not been followed exactly,
different values may appear.
39. Select
Clear Given.
The display should show “Total Given 0 mL”.
40. Press
D
to select
PCA Doses Given.
41. Press
Clear Doses.
The display should show “PCA Doses Given 0, PCA Doses Attempted 0”.
☐
Air Detector Test
Equipment needed
• 1 CADD
™
medication cassette reservoir or 1 CADD
®
administration set
• 1 primed CADD
™
medication cassette reservoir containing fluid and 1 primed CADD
®
extension
set with anti-siphon valve, or
1 primed CADD
®
administration set with anti-siphon valve
• 1 CADD® key
Procedure
1. From the Tasks menu, select
Adjust Admin Settings,
and then select
Alarms
.
Turn the air
detector on.
2. Go back to the Home screen, and then program the pump with the following parameters:
Continuous Rate
PCA Dose
PCA Lockout
Reservoir
Volume
0.0 mL/hr
1.0 mL
1 min
10.0 mL
3. Attach, latch, and lock an
empty
CADD
™
medication cassette reservoir or CADD
®
administration set to the pump.
4. When the screen displays “Reset reservoir volume to 10 mL?”, select
Yes.
5. When the screen displays “Prime tubing?”, select
No.
6. When the screen displays “Start pump?”, select
Yes.
7. The pump should sound a high priority alarm and the display should read “Cannot start pump
with air-in line. Prime tubing.” Select
Acknowledge.
8. Remove the CADD
™
medication cassette reservoir or CADD
®
administration set.
9. Attach, latch, and lock a CADD
™
medication cassette reservoir containing fluid and a primed
CADD
®
extension set with anti-siphon valve, or a primed CADD
®
administration set with anti-
siphon valve to the pump. Make certain there is no air in the fluid path.
10. When the screen displays “Prime tubing?”, select
No.
11. When the screen displays “Start pump?”, select
Yes.
12. Deliver a PCA dose. The pump should deliver the dose without an air detection alarm.
Note:
1 minute must have elapsed since the delivery of the last PCA dose.