18
Batteries
Batteries
Four AA 1.5 volt primary (non-rechargeable) alkaline batteries (for example, Duracell
®
PC1500 / MN1500,
IEC LR6) or the CADD
®
-Solis rechargeable battery pack are recommended for use in the CADD
®
-Solis
pump.
Note:
Smiths Medical does not recommend mixing new and used batteries; doing so may affect low
battery alarm times. Always select four new batteries when replacing depleted batteries.
CAUTION: Inspect the AA batteries for damage or wear to the metal or plastic insulation prior to use, or
after the pump has been dropped or hit. Replace the batteries if any damage is noted.
Replacing the Battery Door
If the battery door is removed or needs replacing, simply snap the door onto the bar
that is located on the pump (see picture).
CADD®-Solis Rechargeable Battery Pack
The battery pack is made up of a lithium-ion cell. When fully charged, its capacity is 5.2 Wh.
Each battery pack can sustain a minimum of 500 charge/discharge cycles. Within the operating
temperature range of 2°C to 40°C (36°F to 104°F), the battery pack becomes fully charged in 4 hours or
less.
The battery pack can be recharged using the CADD
®
-Solis AC adapter. It can be plugged directly into the
AC adapter or it recharges in the CADD
®
-Solis pump with an AC adapter attached.
Note:
Periodically inspect the rechargeable battery pack for damage or wear to the metal or plastic
insulation. Discontinue use if any damage is noted.
See the instructions for use supplied with the rechargeable battery pack for more information.
Battery Storage
The CADD
®
-Solis rechargeable battery pack should not be stored in a refrigerator. Recommended storage
conditions are 19°C to 25°C (66°F to 77°F).
Alkaline batteries should not be stored in a refrigerator. Recommended storage conditions are 10°C to
24°C (50°F to 75°F) with no more than 65% non-condensing relative humidity.
Battery power is quickly depleted at temperatures below 10°C (50°F). After 4 years of storage at 21°C
(70°F), an alkaline battery retains approximately 86% of its original capacity. Battery life is shorter if
the battery is stored above room temperature. An alkaline battery stored at 43°C (110°F) discharges to
approximately 80% of its capacity within one year.
Battery Life
Battery life is dependent on the following factors:
•
Programmed delivery rate
•
Operating temperatures
•
Frequency of use and intensity of display backlighting
•
Duration of use of the USB connector
•
Battery storage conditions
•
Battery type and brand
•