14
Pump Overview
System delivery
accuracy
± 6% (nominal). At low infusion rates, this accuracy may not be achieved for
short periods. During the total infusion time, the accuracy averages out.
WARNING:
•
Ensure that the ± 6% system delivery accuracy specification is taken into
account when programming the pump and/or filling the reservoir. Failure to
do so may result in medication in the reservoir becoming depleted sooner
than expected. If the pump is being used to deliver critical or life sustaining
medication, the interruption in the delivery of medication could result in
patient injury or death.
•
System delivery inaccuracies beyond ± 6% may occur as a result of back
pressure or fluid resistance, which depends upon temperature, drug
viscosity, catheter size, extension set tubing (for example, microbore),
in-line components (such as filters and needleless access connectors), and
placing the infusion reservoir and/or pump above or below the level of the
patient. System delivery inaccuracy may result in under or overdelivery of
medication, which could result in patient injury or death.
Using CADD
™
medication cassette
reservoirs
± 6% (nominal) at 15°C to 40°C with no back pressure
•
An additional ± 2.5% change may be seen at ± 100 mmHg (± 1.9 psi).
Characterization of typical use:
•
An additional –4.8% change may be seen with a typical epidural catheter for
use with an 18 GA needle and with a bolus rate of 40 mL/hr.
•
An additional –7.2% change may be seen with a typical epidural catheter for
use with an 18 GA needle and with a bolus rate of 250 mL/hr.
Using CADD
®
administration sets
± 6% (nominal) at 15°C to 40°C with no back pressure.
•
An additional ± 2.5% change may be seen at ± 100 mmHg (± 1.9 psi).
Characterization of typical use:
•
An additional –5.0% change may be seen with a typical epidural catheter for
use with an 18 GA needle and with a bolus rate of 40 mL/hr.
•
An additional –8.5% change may be seen with a typical epidural catheter for
use with an 18 GA needle and with a bolus rate of 250 mL/hr.
Using CADD
®
high volume
administration sets
± 6% (nominal) at 15°C to 40°C with no back pressure
•
An additional ± 5% change may be seen at ± 100 mmHg (± 1.9 psi).
Characterization of typical use:
•
An additional –7.2% change may be seen with a typical epidural catheter for
use with an 18 GA needle and with a bolus rate of 40 mL/hr.
•
An additional –16.2% change may be seen with a typical epidural catheter
for use with an 18 GA needle and with a bolus rate of 500 mL/hr.
Bolus accuracy
specification: ± 6%
Actual test data for bolus accuracy at 0.05 mL:
Average
0.0508 mL
% Error
1.6%
Minimum Error %
–3.0%
Maximum Error %
4.2%
Actual test data for bolus accuracy at 50 mL:
Average
50.77 mL
% Error
1.55%
Minimum Error %
–0.07%
Maximum Error %
2.35%