13
Pump Overview
Maximum time to
occlusion alarm
and
Maximum bolus
volume at occlusion
alarm
The pressure build-up that occurs after an occlusion may cause fluid to
accumulate or be stored in the line. This extra fluid may be delivered as a bolus
when the occlusion is released.
Note:
Values are representative and based on actual test data.
Flow
Rate
(mL/
hr)
Tubing Set
Max. Time to
Occlusion
Max. Bolus
at Occlusion
Raw
Test
Data
(min)
Spec.
(min)
Raw
Test
Data
(mL)
Spec.
(mL)
0.1
CADD
™
medication cassette
reservoir
<
21-7002 with
CADD
®
extension set
<
21-7047
90
≤ 160
0.107
≤ 0.25
CADD
®
administration set
<
21-7091
122
≤ 190
0.139
≤ 0.30
CADD
®
high volume
administration set
<
21-7055
1140
≤ 1200
1.250
≤ 1.40
Flow
Rate
(mL/
hr)
Tubing Set
Max. Time to
Occlusion
Max. Bolus
at Occlusion
Raw
Test
Data
(sec)
Spec.
(sec)
Raw
Test
Data
(mL)
Spec.
(mL)
10
CADD
™
medication cassette
reservoir
<
21-7002 with
CADD
®
extension set
<
21-7047
50
≤ 60
0.121
≤ 0.25
CADD
®
administration set
<
21-7091
58
≤ 150
0.124
≤ 0.30
CADD
®
high volume
administration set
<
21-7055
545
≤ 580
0.1322
≤ 1.40