20
Batteries
Battery Status
Battery
State
CADD®-Solis Pump Status
25% to 100%
No alarm
Low battery
•
Transition to low battery condition
•
Battery low message appears
•
Pump emits 3 beeps every 5 min
•
Low battery warning message appears on pump display
•
Pump is operable
•
LCD backlight flashes for 12 ms during each motor operation
Depleted
battery
•
Transition to depleted battery condition
•
Battery depleted message appears
•
Pump emits a continuous, variable-tone alarm
•
Depleted battery warning message appears on pump display
•
Battery power is too low to operate pump
•
Pump delivery operation stops
Shut down
Pump shuts off due to low operating voltage
Collect Separately
This product contains electrical and electronic components (including batteries) that may contain
materials, which if disposed of with general waste, could be damaging to the environment.
In accordance with Directive 2002/96/EC Waste Electrical and Electronic Equipment, residents of the
European Union must follow specific disposal or recycling instructions for this product. Contact your
local distributor, or visit the following web site for specific instructions:
http://www.smiths-medical.com/recycle/index.html
Non-European Union residents must dispose of or recycle this product (including batteries) in accordance
with the local laws or regulations that apply.
WARNING:
There are potential health hazards associated with improper disposal of batteries,
electronics, and contaminated (used) reservoirs and extension sets. Dispose of used batteries,
reservoirs, extension sets, and other used accessories, or a pump that has reached the end of its
useful life, in an environmentally safe manner, and according to any regulations that may apply.
Содержание CADD Solis 2100
Страница 1: ...2100 2110 Pump software version 3 0 X s ...