19
Batteries
The following tables may be used to predict typical alkaline battery and CADD
®
-Solis rechargeable battery
pack life at different delivery rates. As expected, battery life decreases as the delivery rate increases.
These tables are based on laboratory tests conducted at room temperature, using fresh Duracell
®
alkaline
batteries and CADD
®
-Solis rechargeable battery packs, in CADD
®
-Solis ambulatory infusion pumps with
CADD
®
administration sets.
Alkaline battery
life with screen
backlight intensity
set to 3
These estimates are based on laboratory tests conducted at room temperature
using new batteries (Duracell
®
PC1500 / MN1500, IEC LR6). Actual battery life
varies depending on the battery brand, shelf life, temperature conditions,
delivery rate, and frequency of screen display and backlighting. It is
recommended that new batteries be kept available for replacement.
Continuous Delivery
Delivery Rate
(mL/hr)
Operating Time
(hr)
Volume Delivered
(mL)
0.4
142
56
1
.
139
139
5
.0
124
620
10
.0
113
1130
30
.0
69
2070
Intermittent Bolus Delivery
IB Volume
(mL)
IB Interval
(min)
Maximum
Delivery
Rate
(mL/hr)
Operating
Time
(hrs)
Volume
Delivered
(mL)
5
30
40
90
900
10
30
250
104
2080
10
30
500
88
1760
Rechargeable
battery pack
life with screen
backlight intensity
set to 3
These estimates are based on laboratory tests conducted at room temperature
using a new CADD®-Solis rechargeable battery pack. Actual battery life varies
depending on the temperature conditions, delivery rate, and frequency of
screen display and backlighting. It is recommended that new batteries be kept
available for replacement.
Continuous Delivery
Delivery Rate
(mL/hr)
Operating Time
(hr)
Volume Delivered
(mL)
0.4
74
29
1
.
67
67
5
.0
60
300
10
.0
50
500
30
.0
40
1200
Intermittent Bolus Delivery
IB Volume
(mL)
IB Interval
(min)
Maximum
Delivery
Rate
(mL/hr)
Operating
Time
(hrs)
Volume
Delivered
(mL)
5
30
40
71
710
10
30
250
60
1200
10
30
500
58
1160
Содержание CADD Solis 2100