12
Pump Overview
Pump alarms
•
High priority alarms: Air in line detected, Battery depleted while delivering,
Battery removed while delivering, Battery unusable while delivering,
Disposable attached improperly, Disposable damaged, Disposable
detached while delivering, Disposable locked but not latched, Disposable
type high flow administration set not allowed, Disposable type high
flow administration set required, Disposable type invalid, Downstream
occlusion, Key stuck, Pressure sensor faulty, Pump automatically stopped,
Rechargeable battery end of life, Remote dose cord key stuck, Reservoir
volume empty, Stop mode reminder, Upstream occlusion
•
Medium priority alarms: 19
•
Low priority alarms: 8
•
Informational messages/alerts 23
Delivery rate during
priming
•
Standard volume cassette tubing: approx. 250 mL/hr
•
High flow volume cassette tubing: 500 mL/hr
Alarm disabled
during priming
Air-In Line
Maximum infusion
pressure
1.86 bar
27.0 psi
High pressure alarm
threshold
1.24 bar ± 0.62 bar
18 ± 9 psi
Air detector alarm
Sensitivity:
•
Low: Single bubble > 400 μL
•
High: Single bubble > 150 μL
Accumulated Air:
Greater than 1 mL air over 15 minutes (nominal)
Maximum volume
infused under
single-fault
conditions
•
CADD® administration set: 0.15 mL
•
CADD® high volume administration set: 0.30 mL