11
Pump Overview
Specifications (Nominal)
General Pump Specifications
System definition
CADD®-Solis pump with 1 of the following attached:
•
Medication cassette reservoir and CADD® extension set
•
Medication cassette reservoir with Flow Stop feature and CADD® extension set
•
CADD® administration set
•
CADD® administration set with Flow Stop feature
Classification
CF
J
Class II
K
Used to test the
pump
•
CADD™ medication cassette reservoirs,
<
21-7002 and
<
21-7309
•
CADD® extension sets,
<
21-7047 and
<
21-7046
•
CADD® administration sets,
<
21-7091 and
<
21-7321
•
CADD® high volume administration sets,
<
21-7355 and
<
21-7357
Resolution
•
CADD™ medication cassette reservoir: 0.050 mL per pump stroke nominal
•
CADD® administration set: 0.050 mL per pump stroke nominal
•
CADD® high volume administration set: 0.1 mL per pump stroke nominal
Size
Excluding cassette and accessories:
4.1 cm × 10.2 cm × 12.7 cm
1.6 in × 4 in × 5 in
Weight
Including 4 AA alkaline batteries, excluding other accessories:
595 g
21 oz
System operating
temperature
15°C to 40°C
59°F to 104°F
System storage
and transportation
temperature
–20°C to 60°C
–4°F to 140°F
Moisture protection
Splashproof (
E
) per IEC 60529
Relative humidity
20% to 90% relative humidity, non-condensing
Atmospheric
pressure
70 kPa to 106 kPa
10.2 psi to 15.4 psi
Power sources
•
AC adapter
•
CADD®-Solis rechargeable battery pack
•
Four AA alkaline batteries (for example, Duracell
®
PC1500 / MN1500, IEC LR6)
Charging system
for internal memory
backup battery
The internal memory backup battery uses lithium manganese dioxide
technology. It charges whenever the pump is powered on and has a 10-month
memory capacity once it has been charged for 250 hours at 20°C (68°F).
Battery fallout alarm Alarm sounds for 2 minutes if the pump has been powered up for a minimum
of 4 minutes.
Note
: Alarm enabled while pump is in run mode only.