Annual Inspection and Testing
☐
Motor and “Reservoir Volume is Zero” Alarm Checks
Note:
Conduct the Motor and “Reservoir Volume is Zero” Alarm check, and Stopping/Starting the
Pump and LED Indicator tests in sequence, as shown below.
Equipment needed
• 1 primed CADD
™
medication cassette reservoir containing fluid, and 1 primed CADD
®
extension
set with anti-siphon valve, or
1 primed CADD
®
administration set with anti-siphon valve, or
1 CADD
®
administration set with Flow Stop feature.
• 1 CADD® key
• Timer
• CADD
®
-Solis remote dose cord
Procedure
1. Program the pump as follows.
a. From the home screen, select
Tasks.
b. Under the
Tasks
menu, select
Start New Patient
.
c. For therapy, select
[Program Manually].
d. For units, select
mL.
e.
Confirm the settings are correct, and select
Yes.
f. The screen displays the “Review pump settings” screen. Select
Review.
g.
Program the following parameters.
Continuous Rate
PCA Dose
Reservoir Volume
30.0 mL/hr
0.0 mL
1.0 mL
h. Once each parameter is correct, select
Accept Value
for that setting.
i. When the review is complete, select
Next.
j. The screen displays “Cassette not attached. Attach cassette before starting pump.” Select
Home.
2. Attach a CADD
™
medication cassette reservoir or CADD
®
administration set to the pump.
Latch and lock the cassette.
3. The screen displays “Prime tubing?”. Prime the pump as follows.
a. Select
Yes.
b. Select
Prime
.
c. The screen displays “Reservoir Volume low. Are you sure you want to continue priming?”.
Select
Yes.
4. The pump should begin to prime. Listen to the motor for excessive noise or grinding sounds.
5. The pump should deliver 1.0 mL, and then the display should show “Reservoir Volume is zero.
Pump stopped.” Select
Acknowledge
.
6. The reservoir volume in the upper left corner of the display should show 0 mL. If the display
shows a different value, verify the pump settings and rerun this test.
7. The screen displays “Reset reservoir volume to 1 mL? Reservoir volume is low.” Select
No.
8. The screen then displays “Start pump?”. Select
No.
☐
Stopping and Starting the Pump / LED Indicator Test
9. Press stop/start
9
.
10. The display should show “Reset reservoir volume to 1 mL? Reservoir volume is low.” Select
Yes.
11. The display should show “Start pump?”. Select
Yes.