5 Operation
64 08 137 D3610
D3610.201.02.09.02 11.2017
77
5.6.1.3
Posterior tooth exposures
For posterior tooth exposures use the yellow sensor holder.
1.
Place the yellow localizer ring (B) onto the double-angled guide rod
(C).
2.
Place the yellow sensor holder tab (A) onto the guide rod (C).
3.
Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
4.
For the left upper jaw and right lower jaw: Glue the sensor holder
tab onto the sensor's hygienic protective sleeve. Place the tab in the
center on the sensor. The edge of the tab must lock with the edge
of the sensor, as shown in the diagram.
5.
For the right upper jaw and left lower jaw the sensor holder tab must
be placed in the mirrored position. See the adjacent drawing.
6.
Position sensor in patient's mouth.
7.
Bring the X-ray tube assembly into the correct position and take an
X-ray exposure.
8.
Remove the sensor from the hygienic protective sleeve. For this
follow the instructions in section "Remove the hygienic protective
sleeve from the sensor" [→ 95]. The used sensor holder tab and
hygienic protective sleeve must be disposed of after the
examination.
9.
Clean and sterilize the guide rod and localizer ring.