4.9.2
Pairing using RFID Technology
RFID technology is particularly useful where multiple WiFi interfaces are
being used for different PCs in the dental practice or clinic. Pairing and
disconnection take place via an RFID transmitter which is in the WiFi
interface. It is read by an RFID reader connected to the PC.
Plug RFID reader into the PC
The RFID reader can be connected to any USB port on the PC,
including to monitors and keyboards with a USB port as well as external
hubs. Select a slot that is easily accessible to the user. A brief message
in the Windows information area (Systray) confirms the successful
pairing. As a result the user should be able to see the monitor during
the pairing.
The RFID readers for the XIOS XG WiFi system are from Sirona.
➢ Plug the RFID reader into an easily accessible USB port, e.g. on the
monitor. If necessary, use the USB extension cable supplied for
improved accessibility.
Ä
The driver for the RFID reader installs automatically. A
message appears in the Windows information area (Systray)
after a few seconds to confirm that the driver can now be used.
Enable pairing via RFID technology
The option in the configuration program of the SIDEXIS plug-in for XIOS
XG must be activated in order that the RFID technology can be used.
1.
Start the SIDEXIS Manager.
2.
Start the
"XIOS XG Configuration" configuration tool.
3.
Select the
"Device settings" tab.