4 Installation
Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
32
64 08 137 D3610
D3610.201.02.09.02 11.2017
4
Installation
4.1
Position the PC
Depending on whether the PC is operated inside or outside of the
patient environment (up to 1.5 m around the patient), an additional
protective ground conductor must be provided on the PC housing.
When using the XIOS XG WiFi system, Sirona recommends positioning
the PC outside of the patient environment.
CAUTION
Leakage currents from the PC are transmitted to the X-ray system.
If a PC is not sufficiently grounded, there is a risk of electric shock for
both the patient and the user.
➢ The PC must be connected to a grounded electric outlet when in
operation.
➢ If the PC is operated in the patient environment (up to 1.5 m
around the patient), it must also be fitted with an additional
protective ground conductor.
4.2
Install the PC software
The PC must be in a state of operational readiness before the
installation of the XIOS XG Intraoral System. Make sure that the
hardware and the operating system are properly installed. Refer to the
manuals for your PC and operating system.
The installation of SIDEXIS XG / SIDEXIS 4 is described in the
installation instructions for SIDEXIS XG / SIDEXIS 4.
The SIDEXIS plug-in for the XIOS XG WiFi system must be installed in
addition to SIDEXIS XG / SIDEXIS 4. To do this, proceed as follows:
ü
SIDEXIS XG / SIDEXIS 4 (SIDEXIS XG 2.56 or higher) is installed
on the PC.
1.
Log onto your PC with an administrator account.
2.
Insert the
"Sirona XIOS XG Select/Supreme 1.2 - Installation" CD
into your PC's CD/DVD drive.
3.
The setup normally starts automatically. If the setup does not start,
double-click on the "Autorun.exe" file in the main directory of the
installation CD.
Ä
The setup window opens.