Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
5 Operation
64 08 137 D3610
D3610.201.02.09.02 11.2017
75
5.6.1
Position sensor with single-use sensor holder system
5.6.1.1
Notes on the sensor holder tabs
IMPORTANT
The sensor holder tabs are single use devices and cannot be reused.
IMPORTANT
Adhesive properties may deteriorate over time based on age and
storage conditions.
➢ If more than one year has passed since the manufacture date,
check adhesive strength before use.
Do not use if holder does not securely adhere to sheath.
5.6.1.2
Anterior tooth exposure
For anterior tooth exposures use the blue sensor holder.
1.
Place the blue localizer ring (B) onto the triple-angled guide rod (C).
2.
Place the blue sensor holder tab (A) onto the guide rod (C).
3.
Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
4.
Glue the sensor holder tab onto the sensor's hygienic protective
sleeve. Place the tab in the center of the sensor as shown in the
diagram.