Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
6 Maintenance and inspection
64 08 137 D3610
D3610.201.02.09.02 11.2017
105
3.
Using the screwdriver supplied with the set remove the 2 screws
which fasten the cable to the sensor.
Replace the Elastomer strip
XIOS XG Supreme
XIOS XG Select
1.
For XIOS XG Select sensors:
Remove the existing elastomer strip from the sensor with tweezers.
The blue plastic frame remains on the sensor.
For XIOS XG Supreme sensors:
When a red plastic frame with elastomer strip is used: Remove the
entire plastic frame with the elastomer strip using a suitable tool. It
is replaced with a white elastomer part.
When a white elastomer part is used: Remove the entire elastomer
part using a suitable tool.
IMPORTANT
Use the new parts supplied with the cable for the assembly!
XIOS XG Select
XIOS XG Supreme
2.
For XIOS XG Select sensors:
Insert a new blue Elastomer strip in the slot. Check that it is
positioned correctly by pressing lightly with tweezers. The
Elastomer strip must be placed exactly in the slot so that the sensor
works.
For XIOS XG Supreme sensors:
Place a new white elastomer part on the sensor.
Attach a replacement cable
1.
Place the plug for the sensor cable properly on the sensor. Both
parts must interlock.
Содержание XIOS XG Select
Страница 119: ......