4 Installation
båÖäáëÜ
båÖäáëÜ
CAUTION
Strong bending or improper routing of the optical fiber inside the handpiece
may damage the optical fiber, creating a health hazard for patients, dentists
and assistants.
CAUTION
If you switch on the laser unit and the signal of the red aiming laser is not visi-
ble or if you cannot see the aiming beam during treatment, proceed as follows:
¾
Check the optical fiber as well as the fiber connection for mechanical
damage.
If the optical fiber is damaged, replace it with a new one.
¾
Connect the fiber to the laser.
¾
Check the optical fiber whether the signal of the aimingbeam is visible
¾
If you cannot detect any damage on the optical fiber and the signal of the
aiming laser is not visible with the new optical fiber either, turn off the
laser and contact Sirona Dental Systems GmbH, your local dental
distributor or your authorized service center.
WARNING
Using the laser unit without a properly working aiming beam may cause inju-
ries to operators, assistants and patients.
CAUTION
The optical fiber must be cleaned and disinfected prior to the first treatment.
CAUTION
Remove the protection caps only for operation!
CAUTION
Do not touch the end surfaces of connectors and protect against damage!
CAUTION
Do not kink or pinch the optical fiber as this would cause breaking the optical fi-
ber!
CAUTION
Do not pull on the optical fiber!
CAUTION
Note the maximum permissible bending radius of the optical fiber!
• Short-term (during use): 100 x radius of optical fiber
• Long-term (during storage): 600 x radius of optical fiber
NOTICE
i
Magnet is for recognizing the presence and absence of the fiber.
Inserting the optical fiber into a sterilized handpiece
¾
Loosen the proximal part of the handpiece.