60 90 570 D 3485
36
D 3485.201.04.09.02
09.2007
5 Operating Instructions
parameter value can be increased in 1,000 Hz increments. When you
press the “RIGHT ARROW” key for more than 3s , the value increases at
a faster rate.
6.
Treatment energy
: The laser unit automatically calculates the energy
(in mJ) applied during treatment from the power value and the selected
laser activation time. If no activation time has been preset, the energy will
not be calculated.
WARNING
As soon as you have selected the program, the SIROLaser is in standby
mode. As soon as you press the foot switch or finger switch, the laser unit is
activated.
When you actuate the foot switch or finger switch, the message “LASER ON”
appears. At the same time, 4 yellow LEDs light up, the audible alarm sounds
and the laser starts emitting. When you release the foot switch or finger switch
to interrupt treatment, the laser is deactivated, but remains ready for opera-
tion.
To change the output parameters, press the “ENTER” key. As soon as the
area has been selected, you can change the parameters by pressing the
"RIGHT ARROW" or "LEFT ARROW" key.
When you press the red “LASER STOP” emergency button, laser radiation
(and thus also treatment) is terminated immediately.
To quit the selected menu or save your changes, press the “ENTER” key for
more than 1 second. A message confirming the end of the therapy appears
on the display.
¾
The “LEFT ARROW” key returns you to the main menu.
5.9
“SETUP AREA” submenu
After accessing the “SETUP AREA”, you can scroll through the available
functions with the “ENTER” key. The following functions are available:
FUNCTION 1:
Adjustment of the display contrast.
FUNCTION 2:
Activation or deactivation of the key tones.
FUNCTION 3:
Activation or deactivation of the audible alarm during laser
radiation (Sirona Dental Systems GmbH recommends acti-
vating this function).
FUNCTION 4:
Activation or deactivation of the auto-repeat function for the
laser activation time in single pulse mode.
FUNCTION 5:
Selection of the interface language
FUNCTION 6:
Service program for laser calibration