60 90 570 D 3485
D 3485.201.04.09.02
09.2007
27
4 Installation
båÖäáëÜ
båÖäáëÜ
4.7.4
Adjusting the sheath
The silicone sheath should be 5-10 cm shorter than the optical fiber itself. If
the sheath is too short, the optical fiber may not be fastened properly in the
handpiece. If the sheath is too long, the optical fiber cannot be inserted into
the handpiece. To shorten the sheath to an appropriate length in relation to
the optical fiber, proceed as follows:
¾
Hold the optic fiber approx. 10 cm below the distal end.
¾
Press the sheath and the optical fiber together.
¾
Pull off approximately 10cm of the sheath with your thumbnail.
NOTICE
i
After
s
tripping the optical fiber, bend it back and forth to make sure it has not
been damaged.
4.8
Power supply connection
¾
Connect the power cord to the DC IN socket at the back of the
SIROLaser (see figure 22).
Figure 22: Power supply connection
CAUTION
The SIROLaser may only be operated with the Sinpro MPU50-105 power sup-
ply. Operation with other power supplies may result in failure or destruction of
the laser unit. Using power supplies other than the recommended one results
in invalidation of the approval of the entire unit and loss of the warranty granted
by Sirona Dental Systems GmbH.
WARNING
Using power supplies other than the recommended one may cause the risk of
overheating and malfunction of the laser unit.
Figure 21: Sheath of the optical fiber is
too long
Содержание sirolaser blue
Страница 1: ...k W MVKOMMT b pfoli l f...
Страница 57: ......