60 90 570 D 3485
22
D 3485.201.04.09.02
09.2007
4 Installation
¾
Plug the connector of the finger switch cable into the socket of the Y
connector or into the interlock jumper.
4.7
Connecting the optical fiber
4.7.1
Field of application
For different dental procedures and indications, two SIROLaser optical fibers
with different diameters are supplied with your SIROLaser: the 200µm fiber
(bend protection colored black) and the 320 µm fiber (bend protection colored
yellow). The user will be prompted by the laser software to select the appro-
priate fiber for each indication.
The SIROLaser optical fibers have a standard SMA-905 connection and can
be used only with the SIROLaser in the spectral range of 970nm
±
15 nm.
If optical fibers from other manufacturers are used, physical properties such
as tensile strength and transmission behavior may vary. For this reason,
Sirona Dental Systems GmbH assumes no liability in such cases.
Use Sirona optical fibers only.
4.7.2
Initial use of an optical fiber
Initial check
¾
Before using an optical fiber for the first time, check its best-before date.
A nonsterile optical fiber can be used maximally 4 years after its month of
manufacture. This information is printed on the product label of the
packaging.
WARNING
If the optical fiber is used after the best-before date, some of its physical
properties, e.g. its load carrying strength and transmission behavior, will
change, thus posing a hazard to the health of the patient, the dentist and the
dental assistant.
¾
Do not use the laser probe if its packaging is damaged or the best-before
month has been exceeded.
¾
After removing the optical fibers from their original packaging, perform a
visual check to make sure that they were not damaged during shipment.
Initial startup
CAUTION
The optical fiber and the take-up reel must be disinfected prior to initial use.
Figure 11: Connection of the finger
switch